Stability of the K4b2 portable metabolic analyser during rest, walking and running

Gespeichert in:
Bibliographische Detailangaben
Deutscher übersetzter Titel:Stabilität des tragbaren K4b2-Stoffwechselanalysegerätes während Erhohlung, Gehen und Laufen
Autor:Howe, Christopher C.F.; Matzko, Richard O.; Piaser, Fabio; Pitsiladis, Yannis P.; Easton, Chris
Erschienen in:Journal of sports sciences
Veröffentlicht:32 (2014), 2, S. 157-163, Lit.
Format: Literatur (SPOLIT)
Publikationstyp: Zeitschriftenartikel
Medienart: Gedruckte Ressource
Sprache:Englisch
ISSN:0264-0414, 1466-447X
DOI:10.1080/02640414.2013.812231
Schlagworte:
Online Zugang:
Erfassungsnummer:PU201407006410
Quelle:BISp

Abstract

This study investigated the stability of the measurement of respiratory variables during rest, walking and running using the K4b2 portable metabolic analyser in ten active males (age 31 ± 11 years; VO2 peak 42.1 ± 2.6 ml · min−1 · kg−1). Following a 10 min rest, participants completed three discontinuous incremental exercise tests on a treadmill while walking (4, 5 and 6 km · h−1) and running (8, 10, 12, 14 and 16 km · h−1, or until volitional exhaustion). Participants completed 3 min of exercise at each speed, followed by a 3 min recovery after each stage above 10 km · h−1. The respiratory variables were measured using either a laboratory-based metabolic cart as a reference method (Oxycon Pro, OP), a K4b2 calibrated immediately before the test (K4b2), or a K4b2 calibrated 1 h before the test (K4b2DEL). Compared to the OP, carbon dioxide production (VCO2) and VE were not different when measured by K4b2 or K4b2DEL. There was no difference in VO2 between OP and K4b2 tests (P = 0.19, mean difference = 38 ml · min−1 and limits of agreement (LOA) = 208 to –285) although K4b2DEL overestimated VO2 (P = 0.05, mean difference = 84 ml · min−1 and LOA = 302 to –469). These data suggest that a drift in measurement accuracy appears to cause the K4b2 to overestimate VO2 in tests lasting longer than 1 h. Verf.-Referat