Reliability of air displacement plethysmography

Gespeichert in:
Bibliographische Detailangaben
Deutscher übersetzter Titel:Reliabilität der Luftverdrängungsplethysmographie
Autor:Anderson, Dawn E.
Erschienen in:Journal of strength and conditioning research
Veröffentlicht:21 (2007), 1, 169-172, Lit.
Format: Literatur (SPOLIT)
Publikationstyp: Zeitschriftenartikel
Medienart: Gedruckte Ressource
Sprache:Englisch
ISSN:1064-8011, 1533-4287
Schlagworte:
Online Zugang:
Erfassungsnummer:PU201006004950
Quelle:BISp

Abstract

The purpose of this study was to examine the reliability of an air displacement plethysmography device (BOD POD) over trials performed on 3 different days. Subjects consisted of 24 healthy adults (8 men, 16 women), ages 18-38 years, with body weights 46.8-93.6 kg, body mass indexes of 19.1-30.1 kg[middle dot]m-2, and percentage body fats (BF) of 7.9-43.1%. Two estimates of BF were performed on 3 days. Paired t-tests revealed no significant within-day differences in body volume (BV), thoracic gas volume (VTG), body density (BD), and BF. Correlations between the two VTG measures on a day were r = 0.86 for day 1, r = 0.93 for day 2, and r = 0.96 for day 3. BF estimates within a day had high correlations of r = 0.98. Significant differences were found between days for measures of BV, VTG, BD, and BF. These results indicate a high reliability for within-day estimates of BF and significant differences in between-day estimates of BF using air displacement plethysmography. Reliability of BF may be increased by requiring subjects to practice the procedure for VTG measurement. Verf.-Referat