Genucom knee analysis system: reproducibility and database development

Gespeichert in:
Bibliographische Detailangaben
Deutscher übersetzter Titel:Genucom Kniegelenkanalysesystem: Reproduzierbarkeit und Datenbank-Entwicklung
Autor:Highgenboten, Carl L.; Jackson, Allen; Meske, Neil B.
Erschienen in:Medicine and science in sports and exercise
Veröffentlicht:22 (1990), 5, S. 713-717, Lit.
Format: Literatur (SPOLIT)
Publikationstyp: Zeitschriftenartikel
Medienart: Gedruckte Ressource
Sprache:Englisch
ISSN:0195-9131, 1530-0315
Schlagworte:
Online Zugang:
Erfassungsnummer:PU199302047044
Quelle:BISp

Abstract

The Genucom is a relatively new computerized system that performs a multidimensional analysis of knee laxity. Therefore, the purposes of the present investigation were twofold: 1) to estimate the reproducibility of the testing system in its ability to obtain similar test values on the same test subject within a back-to-back two-test protocol and 2) to develop a database of normal values as a standard against which values of suspected ligament injury can be compared to reliably determine whether, in fact, ligament injury has occurred. In the reproducibility phase of the study, 20 subjects were tested in two trials each of eight tests, yielding 20 separate test values per trial, accounting for each angle of flexion and plane of force tested. Both legs were tested, providing an N of 40. Test-retest correlations of the two trials and dependent t-tests indicated that the Genucom provided generally stable and reproducible results. In the normal value database phase of the study, 218 different subjects were given one trial of the same eight tests. Testing of both legs provided an N > 400. Means and standard deviations were calculated for each test. Verf.-Referat