The accuracy and reliability of commercial heart rate monitors

Gespeichert in:
Bibliographische Detailangaben
Deutscher übersetzter Titel:Die Genauigkeit und Verlaesslichkeit im Handel erhaeltlicher Herzfrequenzmessgeraete
Autor:Burke, M.J.; Whelan, M.V.
Erschienen in:British journal of sports medicine
Veröffentlicht:21 (1987), 1, S. 29-32, Lit.
Format: Literatur (SPOLIT)
Publikationstyp: Zeitschriftenartikel
Medienart: Gedruckte Ressource
Sprache:Englisch
ISSN:0306-3674, 1473-0480
Schlagworte:
Online Zugang:
Erfassungsnummer:PU198705029697
Quelle:BISp

Abstract

In recent years increasing emphasis has been placed on the benefits of exercise to normal healthy adults as well as to those recovering from heart attacks and other illnesses. As an aid to exercise training many portable heart-rate monitors have been placed on the market but little appears to have been done to assess the performance of these machines. The authors have examined the performance of four such monitors both under bench-test and practical exercising conditions. When tested on the bench using electronic equipment, the machines rarely exhibited errors exceeding 2-3 bpm over a measurement range of 30-240 bpm. However, when tested with subjects walking or jogging at low speeds on a treadmill, typically 20-70 of the readings given by the machines had errors of greater than 20 bpm. In some cases over 50 of readings had errors exceeding 50 bpm. Verf.-Referat