Immunoreactive LH in long-term frozen human urine samples

Gespeichert in:
Bibliographische Detailangaben
Deutscher übersetzter Titel:Immunoreaktives LH in langzeitgefrorenen menschlichen Urinproben
Autor:Singh, Gurmeet Kaur Surindar; Jimenez, Mark; Newman, Ron; Handelsman, David J.
Erschienen in:Drug testing and analysis
Veröffentlicht:6 (2014), 3-4, S. 336-341, Lit.
Format: Literatur (SPOLIT)
Publikationstyp: Zeitschriftenartikel
Medienart: Gedruckte Ressource Elektronische Ressource (online)
Sprache:Englisch
ISSN:1942-7603, 1942-7611
DOI:10.1002/dta.1481
Schlagworte:
Online Zugang:
Erfassungsnummer:PU201406005453
Quelle:BISp

Abstract

Urine provides a convenient non-invasive alternative to blood sampling for measurement of certain hormones. Urinary luteinizing hormone (LH) measurements have been used for endocrinology research and anti-doping testing. However, the commercially available LH immunoassays are developed and validated for human blood samples but not urine so that LH assays intended for use with urine samples need thorough validation. Therefore, the present study evaluated the measurement of urinary LH immunoreactivity using previously validated immunofluorometric (IF) and immunochemiluminometric (ICL) LH assays after prolonged frozen storage. LH was measured in serial urine samples following administration of a single injection of one of two doses of recombinant human chorionic hormone (rhCG) with assays run at the end of study (2008) and again after four years of frozen (-20 °C) storage where samples were stored without adding preservatives. The ICL assay showed quantitatively reproducible LH measurements after prolonged -20 °C storage. However, the IF immunoassay gave consistently lower LH levels relative to ICL (2008) with a further proportionate reduction after four years of sample storage (2012). Yet, both the assays displayed similar patterns of the time-course of urine LH measurement both before and after four years of frozen storage. In conclusion, we found that both immunoassays are suitable for urinary LH measurements with ICL assay being more robust for quantitative urinary LH measurement such as for anti-doping purposes, whereas the IF could be applicable for research studies where urine LH levels are compared within-study but not in absolute terms. Verf.-Referat