Clinical evaluation of a new noninvasive ankle arthrometer

Gespeichert in:
Bibliographische Detailangaben
Deutscher übersetzter Titel:Klinische Evaluierung eines neuen nicht-inavasiven Gelenk-Arthrometers
Autor:Nauck, Tanja; Lohrer, Heinz; Gollhofer, Albert
Erschienen in:The physician and sportsmedicine
Veröffentlicht:38 (2010), 2, S. 55-61, Lit.
Format: Literatur (SPOLIT)
Publikationstyp: Zeitschriftenartikel
Medienart: Gedruckte Ressource Elektronische Ressource (online)
Sprache:Englisch
ISSN:0091-3847, 2326-3660
DOI:10.3810/psm.2010.06.1783
Schlagworte:
Online Zugang:
Erfassungsnummer:PU201104003711
Quelle:BISp

Abstract des Autors

Introduction: A nonradiographic arthrometer was developed to objectively quantify anterior talar drawer instability in stable and unstable ankles. Diagnostic validity of this device was previously demonstrated in a cadaver study. The aim of the present study was to validate the ankle arthrometer in an in vivo setting. Methods: Twenty-three subjects participated in the study. An orthopedic surgeon first performed a manual anterior talar drawer test to classify the subjects' ankles as stable or unstable. The subjects were then evaluated using the ankle arthrometer, and filled out a validated self-reported questionnaire (German version of the Foot and Ankle Ability Measure [FAAM-G]). Ankle stiffness was calculated from the low linear region (40-60 N) of the load deformation curves obtained from the ankle arthrometer. Reliability testing of these stiffness values was done based on load deformation curves, with 150 and 200 N maximum anterior drawer loads applied in the ankle arthrometer. Results: Using the manual anterior drawer test, 16 ankles were classified as stable and 7 were classified as unstable. Arthrometer stiffness analysis differentiated stable from unstable ankles (P = 0.00 and P = 0.01, respectively). Test-retest demonstrated an accurate reliability (intraclass correlation coefficient = 0.80). A significant correlation was found between both FAAM-G subscales and the arthrometer stiffness values (r = 0.43 and 0.54; P = 0.04 and 0.01). Discussion Subjects with and without mechanical ankle instability could be differentiated by ankle arthrometer stiffness analysis and the FAAM-G questionnaire results. This nonradiographic device may be relevant for screening athletes at risk for ankle injuries, for clinical follow-up studies, and implementing preventive strategies. Conclusion: Validity and reliability of the new ankle arthrometer is demonstrated in a small cohort in an in vivo setting. Verf.-Referat