Validation of a personal fluid loss monitor

Gespeichert in:
Bibliographische Detailangaben
Deutscher übersetzter Titel:Validierung eines individualisierten Flüssigkeitsverlust-Überwachungsgerätes
Autor:Wickwire, J.; Bishop, P.A.; Green, J.M.; Richardson, M.T.; Lomax, R.G.; Casaru, C.; Jones, E.; Curtner-Smith, Matthew D.
Erschienen in:International journal of sports medicine
Veröffentlicht:29 (2008), 2, S. 139-144, Lit.
Format: Literatur (SPOLIT)
Publikationstyp: Zeitschriftenartikel
Medienart: Gedruckte Ressource Elektronische Ressource (online)
Sprache:Englisch
ISSN:0172-4622, 1439-3964
DOI:10.1055/s-2007-965319
Schlagworte:
Online Zugang:
Erfassungsnummer:PU200808002374
Quelle:BISp

Abstract

Dehydration raises heat injury risk and reduces performance [[2], [5], [6]]. The purpose was to validate the Hydra-Alert Jr (Acumen). The Hydra-Alert was tested in two exercise/clothing conditions. Participants wore it while wearing exercise clothing and exercising at a self-selected intensity (n = 8). Others wore the Hydra-Alert while wearing a ballistic-vest and performing an industrial-protocol (n = 8). For each condition, the Hydra-Alert was tested on two occasions (T1 and T2). The Hydra-Alert was tested against nude weight loss for both conditions. The Hydra-Alert had low test-retest reliability for both conditions (average absolute value of the error between Hydra-Alert outputs of T1 and T2 = 0.08 ± 0.08 percentage points). With exercise-clothing, the Hydra-Alert evidenced low-moderate correlations between percent nude weight loss and Hydra-Alert output at 20 min (r = 0.59-T1, p = 0.13; r = 0.12-T2, p = 0.78), at 40 min (r = 0.93-T1, p = 0.001; r = 0.63-T2, p = 0.10), and at ∼2% weight loss (r = 0.21-T1 and T2, p = 0.61 and 0.62, respectively). The correlation at 40 min during T1 fell during T2 suggesting the Hydra-Alert was inconsistent. When wearing a ballistic-vest, the Hydra-Alert had poor validity (T1: r = -0.29 [p = 0.48] for weight loss vs. monitor; T2: r = 0.11 [p = 0.80]). At the higher levels of dehydration (∼2%), the Hydra-Alert error was so high as to render its readings of little value. In some cases, the Hydra-Alert could lead to a false level of security if dehydrated. Therefore, the Hydra-Alert is of little use for those who want to measure their fluid loss while exercising in the heat. Verf.-Referat