Validity of a swimming snorkel for metabolic testing

Gespeichert in:
Bibliographische Detailangaben
Deutscher übersetzter Titel:Die Validität des Schwimmschnorchels für metabolische Tests
Autor:Rodriguez, F.A.; Keskinen, K.L.; Kusch, M.; Hoffmann, Uwe
Erschienen in:International journal of sports medicine
Veröffentlicht:29 (2008), 2, S. 120-128, Lit.
Format: Literatur (SPOLIT)
Publikationstyp: Zeitschriftenartikel
Medienart: Gedruckte Ressource Elektronische Ressource (online)
Sprache:Englisch
ISSN:0172-4622, 1439-3964
DOI:10.1055/s-2007-964973
Schlagworte:
Online Zugang:
Erfassungsnummer:PU200808002366
Quelle:BISp

Abstract

Two models of a swimming snorkel connected to a portable metabolic cart (Cosmed K4 b2, Rome, Italy) were assessed using a gas exchange simulation system. Four standardized testing protocols were designed to mimic different swimming conditions and were performed similarly in three conditions so that both snorkels could be compared to measured values obtained by connecting the simulator directly with the gas analyzer. Simulated and measured values were highly correlated (R2 = 0.891 to 0.998) and in good agreement, with only a small overestimation of expiratory tidal volume (4%, p = 0.005), not large enough to significantly affect the accuracy of ventilation or gas exchange parameters. Values measured using both swimming snorkels also highly correlated with simulated values, particularly for the ventilatory and primary gas exchange variables (R2 = 0.996 and 0.998 in both models for VO2 and VCO2, respectively). A moderate overestimation of FEO2 was observed in both models (2.65% and 2.48% relative, p = 0.03) and attributed to minimal mixing of inspiratory and expiratory gases, although not affecting VO2 measurements. We conclude that both snorkels are valid devices for measuring pulmonary breath-by-breath gas exchange parameters in connection with the K4 b2 across a wide physiological range. Verf.-Referat