Validity and reliability of the Cosmed K2 instrument

Gespeichert in:
Bibliographische Detailangaben
Deutscher übersetzter Titel:Validitaet und Reliabilitaet des Cosmed K2-Geraets
Autor:Lucia, A.; Fleck, S.J.; Gotshall, R.W.; Kearney, J.T.
Erschienen in:International journal of sports medicine
Veröffentlicht:14 (1993), 7, S. 380-386, Lit.
Format: Literatur (SPOLIT)
Publikationstyp: Zeitschriftenartikel
Medienart: Gedruckte Ressource Elektronische Ressource (online)
Sprache:Englisch
ISSN:0172-4622, 1439-3964
DOI:10.1055/s-2007-1021196
Schlagworte:
Online Zugang:
Erfassungsnummer:PU199311068685
Quelle:BISp

Abstract

The purpose of this study was to measure the validity and reliability of a portable telemetry system (Cosmed K2) for measurement of oxygen consumption (VO2) during a laboratory exercise test at submaximum and maximum intensities. Twenty subjects performed three testing sessions on a treadmill on three consecutive days. Each testing session consisted of a submaximal and a maximal exercise test. On one day, VO2 was determined with the conventional Douglas bag method, and on the other two days, VO2 was determined using the Cosmed K2 system. Utilization of the Douglas bag method or the Cosmed K2 was randomly assigned. The results indicated that the Cosmed K2 instrument is reliable and valied (p>0.05) for measurement of VO2 (l/min) during laboratory exercise testing at submaximal and maximum intensities. The assumption made by the K2 apparatus off a constant respiratory gas exchange ratio (R) of 1.00 did not have a significant influence on VO22 measurements. Verf.-Referat