Monitoring dynamic anterior compartment pressures during exercise

Gespeichert in:
Bibliographische Detailangaben
Deutscher übersetzter Titel:Aufzeichnung des dynamischen Druckes des vorderen Kompartments waehrend koerperlicher Belastung
Autor:McDermott, A.G.P.; Marble, A.E.; Yabsley, R.H.; Phillips, B.
Erschienen in:The American journal of sports medicine
Veröffentlicht:10 (1982), 2, S. 83-89, Lit.
Format: Literatur (SPOLIT)
Publikationstyp: Zeitschriftenartikel
Medienart: Gedruckte Ressource
Sprache:Englisch
ISSN:0363-5465, 1552-3365
Schlagworte:
Online Zugang:
Erfassungsnummer:PU198301017596
Quelle:BISp

Abstract

This report presents the early results of a new technique, using a solid-state transducer intracompartmental (STIC) catheter, developed to measure dynamic pressure changes in the anterior compartment of the leg in patients with exercise-induced anterior compartment syndrome. Nine male volunteers (15 limbs) with a history of exercise-induced anterior compartment pain were studied and compared to eleven asymptomatic male controls matched for age, weight and height . After catheter insertion, anterior compartment pressure were recorded with the subjects walking and running on a treadmill and performing several static maneuvers before and after the exercise period. The most statistically significant value between the two groups was the difference in mean pressure during running. Presently, all mean running pressures greater than 85 mm Hg are considered abnormal. This new technique offers the investigator the opportunity to measuredynamic intracompartmental pressures in symptomatic patients and provides an objective measurement for selection of patients for fasciotomy. Verf.-Referat